Thirty-minute AOBP showed good discrimination for elevated ambulatory blood pressure but systematically overestimated BP relative to ABPM; it may serve as a practical screening tool, although normal AOBP values cannot reliably exclude ambulatory HTN and confirmation with 24-hour ABPM remains necessary.
Key Findings
Results
Thirty-minute AOBP readings were systematically higher than 24-hour and daytime ABPM readings.
Mean differences versus 24-hour ABPM were +6.59 mmHg systolic and +7.27 mmHg diastolic.
Mean differences versus daytime ABPM were +3.73 mmHg systolic and +5.16 mmHg diastolic.
All differences had p < 0.001.
The study enrolled 79 adults undergoing ABPM, with AOBP automated readings every 5 minutes over 30 minutes and the mean of the final six measurements analyzed.
Results
Bland-Altman analysis showed acceptable group-level agreement between AOBP and ABPM but wide limits of agreement.
Agreement was assessed using paired comparisons and Bland-Altman analysis.
Wide limits of agreement indicate substantial individual-level variability between the methods.
Results
Thirty-minute AOBP had high sensitivity but variable, often low, specificity for detecting ambulatory hypertension.
Sensitivity ranged from 84% to 96%.
Specificity ranged from 28% to 79%.
Normal AOBP values cannot reliably exclude ambulatory hypertension.
Results
ROC analysis showed good discrimination of AOBP for elevated ambulatory BP.
AUC ranged from 0.77 to 0.87.
Results
The AOBP cut-off with the highest agreement with overall ABPM was ≥140/84 mmHg.
Cohen's kappa at this cut-off was κ = 0.624.
Cut-off values were identified through ROC analysis and agreement statistics.
Conclusions
The authors conclude that 30-minute AOBP may serve as a screening tool but does not replace 24-hour ABPM for confirmation.
AOBP systematically overestimated BP relative to ABPM.
Confirmation with 24-hour ABPM remains necessary.
What This Means
This study compared a 30-minute automated office blood pressure test, in which a machine takes a reading every 5 minutes without staff in the room, to the standard 24-hour ambulatory monitor worn during daily life. The 79 adults in the study had both tests. The office readings were consistently higher than the 24-hour readings, by about 6-7 mmHg on average, and by somewhat less compared with daytime readings.
The office test was good at picking up people with high ambulatory blood pressure, with sensitivity of 84-96%. It was less good at correctly identifying people without hypertension, with specificity of 28-79%. Agreement was best at a threshold of 140/84 mmHg. Individual differences between the two methods were large.
This research suggests the 30-minute office test could be a practical screening option where 24-hour monitoring is costly or hard to get. However, a normal office result cannot reliably rule out hypertension, and the 24-hour monitor is still needed to confirm. The study was small, at 79 people, so the findings should be confirmed in larger groups.
Bui A, Nguyen Q, Bui A, Ha M, Nguyen N, Nguyen T, et al.. (2026). Compare 30-minute automated office blood pressure vs 24-hour ambulatory blood pressure monitoring: A prospective study.. PloS one. https://doi.org/10.1371/journal.pone.0360957