Sleep

Performance of three consumer sleep-tracking devices compared with Actigraphy and Polysomnography.

TL;DR

Apple Watch Series 7, Fitbit Charge 5, and Polar Vantage M2 demonstrated parameter-specific bias and limited agreement with PSG and actigraphy, suggesting these devices may be appropriate for longitudinal self-monitoring rather than clinical-grade assessment of sleep architecture.

Key Findings

Fitbit Charge 5 significantly overestimated light sleep compared with PSG.

  • Fitbit overestimated light sleep by +57.7 ± 80.5 minutes (P = .009)
  • Sample size for Fitbit group: n = 20
  • Agreement with PSG was generally weak with wide limits of agreement (LOA)
  • No significant differences were reported for Fitbit on other sleep parameters (TIB, TST, REM, deep sleep)

Apple Watch Series 7 significantly underestimated light sleep and overestimated REM sleep compared with PSG.

  • Apple Watch underestimated light sleep by -73.8 ± 61.6 minutes (P < .001)
  • Apple Watch overestimated REM sleep by +30.3 ± 36.3 minutes (P = .003)
  • Sample size for Apple Watch group: n = 19
  • Despite these biases, Apple Watch received the highest overall ranking among the three devices

Polar Vantage M2 showed a significant positive REM bias compared with PSG but no significant differences for other sleep parameters.

  • Polar overestimated REM sleep by +32.3 ± 43.6 minutes (P = .026)
  • No significant differences were found for TIB, TST, light sleep, or deep sleep
  • Sample size for Polar group: n = 15
  • Polar received the lowest overall ranking among the three devices

Agreement between all three consumer wearables and PSG was generally weak with wide limits of agreement across sleep parameters.

  • Agreement was assessed using intraclass correlation coefficients (ICCs) and Bland-Altman analyses
  • Wide LOA were observed across devices for primary outcomes including TIB, TST, REM, light sleep, and deep sleep
  • Overall device ranking favored Apple Watch, followed by Fitbit, then Polar
  • A research-grade actigraph was worn concurrently for 7 days as an ambulatory reference alongside home PSG

The study was a prospective observational design conducted in healthy adults aged 18–45 years with concurrent home PSG and actigraphy as reference standards.

  • Study conducted at King Abdulaziz University Hospital
  • Total participants: Fitbit n = 20, Apple Watch n = 19, Polar n = 15
  • Participants wore a research-grade actigraph concurrently for 7 days
  • Home polysomnography was performed during a home visit as the primary reference standard
  • Participants were healthy adults; clinical populations were not included

Consumer sleep-tracking devices were deemed more appropriate for longitudinal self-monitoring than for clinical-grade assessment of sleep architecture.

  • All three devices showed parameter-specific biases relative to PSG
  • Limited agreement with both PSG and actigraphy was observed across devices
  • Authors concluded devices may suit self-monitoring purposes rather than clinical diagnostics
  • No device demonstrated clinically acceptable agreement with PSG for sleep stage classification

What This Means

This research suggests that three popular consumer wearables — the Apple Watch Series 7, Fitbit Charge 5, and Polar Vantage M2 — do not accurately measure sleep stages when compared against the medical gold standard (polysomnography, or PSG) and a research-grade activity monitor. Specifically, Fitbit tended to overestimate light sleep by nearly an hour, Apple Watch underestimated light sleep and overestimated REM sleep, and Polar overestimated REM sleep. The agreement between all devices and PSG was generally weak, meaning the numbers these gadgets report can differ substantially from what clinical sleep testing would show. The study involved 54 healthy adults aged 18 to 45 who wore one of the three devices alongside a professional actigraphy monitor while also undergoing home sleep testing. Researchers used standard statistical methods (Bland-Altman plots and intraclass correlation coefficients) to measure both the size and consistency of the differences between devices and PSG. Among the three, Apple Watch performed best overall, followed by Fitbit and then Polar, though none performed well enough to be considered clinically reliable for measuring sleep architecture. This research suggests that while these wearables may be useful for tracking general sleep patterns and trends over time, they should not be relied upon for diagnosing sleep disorders or making clinical decisions about sleep health. The specific biases each device shows — such as overestimating certain sleep stages — mean that the sleep stage breakdowns they provide can be misleading when compared to what a sleep specialist would measure in a clinical setting.

Check Your Own Numbers

Upload your bloodwork. We'll cross-reference your results against this study and 4,700 others.

Upload Your Labs

Have a question about this study?

Citation

Alhejaili F, Alshumrani R, Alamoudi W, Wali S, Bamagoos A, Wazira L, et al.. (2026). Performance of three consumer sleep-tracking devices compared with Actigraphy and Polysomnography.. Sleep medicine. https://doi.org/10.1016/j.sleep.2026.109059